Jan. 27, 2020 |
|
|
Nov. 19, 2025 |
|
|
jRCTs052190096 |
Postoperative endovascular intervention for false lumen for aortic dissection (FL intervention) |
|
False lumen intervention (FL intervention) |
|
Mar. 31, 2024 |
|
8 |
|
Median age was 76 (71-78) years, 5 (62.5%) males and 3 (37.5%) females. Six patients (75%) had Stanford type A dissection and two (25%) had type B. Median systolic blood pressure was 124 mm Hg (118-131), median pulse rate was 62 beats/min (60-66), median hemoglobin level was 13.2 g/dl (12.2-13.8), median creatinine was 0.95 mg/dl (0.82-1.21), median eGFR was 52 mL/min/1.73m^2 (46.6-60.9) and median FDP was 27micro-g/mL (13.8-38.8). The median maximum diameter of the aortic aneurysm was 57 mm (51-62), and the patency of the false lumen was partially thrombosed in 7 patients (87.5%) and patent in 1 patient (12.5%). |
|
Enrollment was low compared to the originally anticipated pace, with only 8 patients enrolled. All 8 patients completed endovascular treatment as specified in the protocol; 2 deaths occurred during the 2-year follow-up, but were not related to the study device and treatment. |
|
During the study period, cerebral infarction in 1, subarachnoid hemorrhage in 1, and aspiration pneumonia in 1 patient occurred. These complications were not related to the study device or treatment. |
|
In this study, the primary endpoint was the aortic aneurysm diameter by CT 12 months after treatment with the study device, and whether endovascular treatment with the study device resulted in a reduction in aortic aneurysm diameter was tested. Results showed a median shrinkage of 14% (4.4-20%) in the diameter at 12 months. The patients were also followed up at 24 months with a good median shrinkage of 15.4% (0-27%). |
|
In patients with enlarged false lumen after endovascular repair of graft replacement for aortic dissection, we confirmed the safety and efficacy of using a small-diameter stent graft insertion to close the entry using a test device to prevent the sac enlargement or the false lumen shrinkage. |
|
Nov. 18, 2025 |
No |
|
None |
|
https://jrct.mhlw.go.jp/latest-detail/jRCTs052190096 |
Matsuda Hitoshi |
||
National Cerebral and Cardiovascular Center |
||
6-1 Kishibeshinmachi, Suita, Osaka |
||
+81-6-6170-1070 |
||
hitmat@mist.ocn.ne.jp |
||
Seike Yoshimasa |
||
National Cerebral and Cardiovascular Center |
||
6-1 Kishibeshinmachi, Suita, Osaka |
||
+81-6-6170-1070 |
||
ys46421g@ncvc.go.jp |
Complete |
Jan. 27, 2020 |
||
| Mar. 11, 2020 | ||
| 15 | ||
Interventional |
||
single arm study |
||
open(masking not used) |
||
no treatment control/standard of care control |
||
single assignment |
||
treatment purpose |
||
Patients who had past history of undergoing the surgery for aortic dissection. |
||
Patients who had systemic infection that may increase the risk of device-related infection. |
||
| 20age old over | ||
| No limit | ||
Both |
||
Dissecting aneurysm of the aorta |
||
Entry closure located at the orifice including the neck vessels, visceral arteries, and the iliac arteries. |
||
Stanford classification, Aortic remodeling |
||
Entry closure |
||
I710 |
||
Aortic diameter on CT at 12 months after intervention. |
||
Occlusive condition of the false lumen 12 months after intervention. |
||
| Kobe University Clinical Research Ethical Committee | |
| 7-5-2 Kusunokicho, Chuo-ku, Kobe, Hyogo | |
+81-78-382-6669 |
|
| cerb@med.kobe-u.ac.jp | |
| Approval | |
Jan. 20, 2020 |
Japan |