Validity of the abnormal gait evaluation for stroke patients using the welwalk WW-2000S (VAGEWW) |
|
Abnormal gait evaluation using the welwalk WW-2000S (AGEWW) |
|
Mar. 04, 2025 |
|
34 |
|
Healthy adults: age 16 or over, height between 140cm and 190cm, weight between 40kg and 95kg. Cerebrovascular disease patients: age 16 or over, height between 140cm and 190cm, weight between 40kg and 95kg, who are able to walk using the Welwalk with supervision or independently. |
|
Seventeen healthy adults and 17 individuals with cerebrovascular disorders were included in this study. |
|
As a safety analysis, we counted the number of adverse events that could not deny a causal relationship with the gait training using the Welwalk WW-2000S, and the number of malfunctions with the Welwalk WW-2000S. As a result, there were no adverse events, and there were no malfunctions with the Welwalk WW-2000S. |
|
As for the primary outcome, the abnormal gait pattern index values calculated using the three-dimensional motion analysis system and the Welwalk WW-2000S gait analysis system were shown to have moderate to very high correlations in experiments using simulated walking by healthy adults. The results of experiments on patients also showed the same trend as the results of experiments on healthy adults. As a secondary outcome, there was a high correlation between the scores of the observational assessment for abnormal gait patterns and the abnormal gait pattern index values calculated by the Welwalk WW-2000S gait analysis system in the experiment using simulated walking by healthy people. Furthermore, the Welwalk WW-2000S gait analysis system showed high sensitivity and specificity to detect abnormal gait patterns that could lead to falls in the simulated walking by healthy adults. |
|
The purpose of this study was to assess the validity for the abnormal gait pattern index values calculated by the Welwalk WW-2000S gait analysis system. We confirmed that there was a high correlation between the results of the index values calculated by the Welwalk WW-2000S gait analysis system and the results of index values calculated by the three-dimensional motion analysis system or the scores by observational assessment. |
|
July. 01, 2025 |
Yes |
|
If the joint research facility, Toyota Motor Corporation, asks us to provide the results of the study, we will provide the anonymized information of the participants and not the corresponding tables. |
|
https://jrct.mhlw.go.jp/latest-detail/jRCTs042190109 |
Otaka Yohei |
||
Fujita Health University Hospital |
||
1-98 Dengakugakubo, Kutsukake, Toyoake, Aichi 470-1192, Japan |
||
+81-562-93-2167 |
||
rehabmed@fujita-hu.ac.jp |
||
Hirano Satoshi |
||
Fujita Health University Hospital |
||
1-98 Dengakugakubo, Kutsukake, Toyoake, Aichi 470-1192, Japan |
||
+81-562-93-2167 |
||
sshirano@fujita-hu.ac.jp |
Complete |
Jan. 22, 2020 |
||
| 60 | ||
Interventional |
||
non-randomized controlled trial |
||
open(masking not used) |
||
active control |
||
parallel assignment |
||
screening |
||
Healthy and patient with a cerbulovascular disease |
||
Unstable vital sign |
||
| 16age old over | ||
| No limit | ||
Both |
||
Cerbulovascular disease |
||
Healthy: walking simulating the abnormal gait pattern |
||
The index value of abnormal gait pattern |
||
Sensitivity, specificity, time factor for gait cycle |
||
| TOYOTA MOTOR CORPORATION | |
| Not applicable |
| Fujita Health University Certified Clinical Research Review Board | |
| 1-98 Dengakugakubo, Kutsukake-cho, Toyoake, Japan, Aichi | |
+81-562-93-2865 |
|
| crb-f@fujita-hu.ac.jp | |
| Approval | |
May. 21, 2020 |
none |