Phase III clinical trial to evaluate the diagnostic accuracy of QFR (Quantitative Flow Ratio) using online measurement

Phase III clinical trial to evaluate the diagnostic accuracy of QFR

Mar. 10, 2022

186

Mean age 70.8, Male 78.9%, Mean pulse rate 70.0 bpm, Mean LV EF 60.2

After 14 days screening period participants went under CAG or PCI in which FFR and QFR were measured. Participants were followed up until the day of discharge.

Adverse events in 18/186 patients (9.7%), 24 cases Mild 8, moderate 16, serious 0

Diagnostic accuracy of QFR in comparison with FFR as gold standard of functional assessment (Full Analysis Set) Sensitivity 90.6, Specificity 89.0, Positive predictive value 86.5, Negative predictive value 92.4 Average time to FFR 11.25 minutes, Average time to QFR 6.34 minutes

QFR showed high diagnostic accuracy in comparison with FFR as gold standard of functional assessment of coronary arteries.

No

https://jrct.mhlw.go.jp/latest-detail/jRCT2032210032

van Schaik-Yamamoto Yuko

M&S

3rd floor, 5-1 Nihonbashi Kabutocho, Chuo-ku, Tokyo

+81-9086023560

YVanschaik@medisimaging.com

van Schaik-Yamamoto Yuko

Medis medical imaging systems

3rd floor, 5-1 Nihonbashi Kabutocho, Chuo-ku, Tokyo

+81-3-6778-2589

YVanschaik@medisimaging.com

Complete

April. 19, 2021

May. 25, 2021
191

Interventional

single arm study

open(masking not used)

uncontrolled control

single assignment

diagnostic purpose

- Patients with stable angina or patients who undergo a secondary assessment of stenosis after AMI - 20 years of age or older
- Patients who can give written consent to the participation

- Patients who had myocardial infarction within 72 hours
- Patients with severe asthma or severe chronic obstructive pulmonary disease
- Patients with severe heart failure

20age old over
No limit

Both

Stable angina or stenosis after AMI

FFR will be measured during coronary angiography or coronary intervention in the patients who gave written consent. QFR will be measured by a certified user for all vessels for which FFR assessment is planed. Patients will be treated according to the results of angiography and FFR. QFR values should not be used to determine treatment strategy.

Evaluate the diagnostic accuracy of QFR with FFR as gold standard, and compare the diagnostic accuracy of QFR with that of anatomical assessment by QCA.

Medis medical imaging systems
Wakayama Medical University IRB
811-1 Kimiidera, Wakayama-shi, Wakayama

+81-73-441-0547

chiken@wakayama-med.ac.jp
Approval

April. 28, 2021

none

History of Changes

No Publication date
4 July. 14, 2022 (this page) Changes
3 Oct. 20, 2021 Detail Changes
2 May. 31, 2021 Detail Changes
1 April. 12, 2021 Detail