Data collection for a breast ultrasonic imaging device |
|
Data collection for a breast ultrasonic imaging device |
|
Sept. 30, 2021 |
|
200 |
|
This study was conducted on women over the age of 20. Two hundred patients were enrolled.One withdrew consent and ring echo screening was performed on 199 patients. 198 patient's data was analyzed. One patient was excluded as an overflow of hot water from the inspection tank meant the inspection could not be completed. The 198 patients had an average age of 49.7, an average height of 157.9 cm, and an average weight of 56.3 kg. |
|
March 23, 2021: First case registration June 5, 2021: Final case registered September 30, 2021: Follow-up completed |
|
No adverse events occurred. There were two equipment failures. One was physical damage,and the other was an error in recovery work from detected anomalies. Neither led to adverse events. |
|
Primary endpoint: In the ring echo images, the incidence of images with artifacts due to lack of penetration inside the breast was 10.0% (95% CI: 9.6-10.4%), and the degree of impact (ratio of artifact area to breast area) was 0.25%. Secondary endpoint: GCNR (The Generalized Contrast-to-Noise Ratio) of the tumors visualized by the ring echo images and diagnosed as malignant masses and benign masses in normal medical care were 0.66 (95% CI: 0.61-0.71) and 0.65 (95% CI: 0.59-0.71). |
|
We obtained registration of 200 women over 20 years old, performed ring echo examinations of 198, and succeeded in collecting breast images.Insufficient penetration was confirmed in 10% of the acquired images, but the degree of influence was as low as 0.25%. Regarding safety, no adverse events were found. |
|
April. 06, 2022 |
No |
|
No |
|
https://jrct.mhlw.go.jp/latest-detail/jRCTs032200416 |
Furusawa Hidemi |
||
Breast surgical oncology clinic, Kawaguchi kogyo general hospital |
||
F1-2 BRANZ Kawaguchi-Motogo 1-18-21 Aoki, Kawaguchi, Saitama |
||
+81-48-229-0707 |
||
hidemi.furusawa@kogyohsp.gr.jp |
||
Furusawa Hidemi |
||
Breast surgical oncology clinic, Kawaguchi kogyo general hospital |
||
F1-2 BRANZ Kawaguchi-Motogo 1-18-21 Aoki, Kawaguchi, Saitama |
||
+81-48-229-0707 |
||
hidemi.furusawa@kogyohsp.gr.jp |
Complete |
Mar. 12, 2021 |
||
| Mar. 23, 2021 | ||
| 200 | ||
Interventional |
||
single arm study |
||
open(masking not used) |
||
uncontrolled control |
||
single assignment |
||
diagnostic purpose |
||
1)Female, Age 20 or older at the time of consent acquisition |
||
1) Unable to get on and off the device by oneself or unable to lay prone on the device for 30 minutes |
||
| 20age old over | ||
| No limit | ||
Female |
||
Breast diseases |
||
Breast imaging using a device |
||
Evaluation of image |
||
1)Evaluation of image |
||